| Stability & Storage | Store at 2–8°C, protected from light, for 6 months. After reconstitution, the calibrator may be aliquoted and stored at −20°C for up to 3 months; avoid repeated freeze-thaw cycles. |
Background
Human interleukin-13 (Human Interleukin 13, IL-13) is a type I cytokine secreted by activated Th2 cells, with a molecular weight of approximately 12 kDa, and is an important member of the Th2 cytokine family. IL-13 exerts multiple biological functions in immune regulation, particularly playing a critical role in B-cell maturation, differentiation, and immunoglobulin class switching. It can upregulate the expression of CD23 and MHC class II molecules on the B-cell surface and promote class switching toward IgE production, thereby serving as a central mediator in allergic immune responses.In addition,IL-13 exerts a significant negative regulatory effect on the monocyte-macrophage system, inhibiting the production of pro-inflammatory cytokines and chemokines, thereby participating in the regulation of inflammatory responses. In the lungs, IL-13 is one of the key mediators that induce allergic airway diseases, capable of inducing airway hyperresponsiveness, goblet cell metaplasia, and excessive mucus secretion. These pathological changes collectively lead to airway obstruction and impaired pulmonary function, forming an important pathophysiological basis for allergic airway diseases such as asthma.Given thatIL-13 plays a crucial role in allergic diseases, fibrotic disorders, and the immune responses associated with certain cancers, making it a focus of research on disease mechanisms, drug efficacy evaluation, and biomarker development. Researchers often assess immune response status, inflammation severity, and the effects of pharmacological interventions by measuring IL-13 expression levels. The Human IL-13 ELISA Kit provides a sensitive, specific, and efficient tool for the quantitative in vitro detection of IL-13 concentrations in samples such as cell culture supernatants, serum, and plasma, and is suitable for basic biomedical research and drug screening applications..
Principle of the test:

This kit employs a homogeneous chemiluminescence assay based on the sandwich ELISA method.Homogeneous Immuno Chemiluminescence Assay, HICA) to detect cytokine concentrations, with simple operation,No washing required.
The detection system includes two types of microspheres:DonorMicrosphere conjugation targeting the target proteinAntibody 1,ReceptorMicrosphere conjugationAntibody to the target protein 2At the time of the reaction, the target protein binds to the antibody and the microbeads, forming an immune complex that brings the two types of microbeads closer together. When the distance between the donor microbead and the receptor microbead is less than 200 nm, the singlet oxygen generated upon light excitation can be transferred to the receptor microbead, triggering chemiluminescence; conversely, if the target protein is absent and the microbeads are too far apart, no signal is produced.
By detecting the intensity of the chemiluminescence signal, quantitative analysis of the target protein can be achieved. This method is characterized by its simple operation, rapid reaction, and high sensitivity.
Components
Name |
Specifications |
Component specifications |
Testing reagentR1 |
250T |
14mul /Pipe ×1 |
1200T |
seventymul /Pipe ×1 |
|
Testing reagentR2 |
250T |
14mul /Pipe ×1 |
1200T |
seventymul /Pipe ×1 |
|
Calibrator |
250T |
30ngLyophilized product ×1 |
1200T |
30ngLyophilized product ×5 |
|
Reagent diluent |
250T |
5.5 mL/Bottle ×1 |
1200T |
27.5 mL/Bottle ×1 |
|
Calibrator diluent |
250T |
1 ml/Pipe ×1 |
1200T |
1 ml/Pipe ×5 |
Note: It is recommended to use a microplate for the plates used.384Shallow holeor384,96Orifice plate, white or gray)
Protocol
I. Sample Requirements
1.To remove impurities from the sample, the sample must be centrifuged prior to analysis.2000 ×g,10minutes).
2.If the measured concentration of a sample exceeds the highest value on the standard curve, it should be appropriately diluted (using negative serum is recommended) and retested to ensure that the results fall within the valid measurement range.
II. Testing Procedure
2.1Preparation of calibration standard gradient samplesAdopt 100muLReconstitute with deionized water.30ng lyophilized calibrator as C12Then dilute the calibrator with the calibrator diluent; calibrator C.1This is the calibrator diluent; the recommended dilution protocol is shown in the table below:
Gradient |
Concentration (pg/mL) |
schoolQualified product(microliters) |
Diluent (μL) |
C12 |
300,000 |
0μL |
100 |
C11 |
100,000 |
30μL C12 |
60 |
C10 |
30,000 |
30μL C11 |
seventy |
C9 |
10000 |
30μL C10 |
60 |
C8 |
3000 |
30μL C9 |
seventy |
C7 |
1000 |
30μL C8 |
60 |
C6 |
300 |
30μL C7 |
seventy |
C5 |
100 |
30μL C6 |
60 |
C4 |
30 |
30μL C5 |
seventy |
C3 |
10 |
30μL C4 |
60 |
C2 |
3 |
30 μL C3 |
seventy |
C1 |
0 |
- |
- |
2.2Testing procedure:
|
Testing process 37°COne-step methodDetection | |
Add sample |
1.take2µL sampleJoin384Shallow-hole plate; 2.willR1andR2Ultrasound 5sAfter vortexing or stirring to mix, combine equal volumes and mix thoroughly, then dilute with the reagent diluent.200Mix and dilute to prepare the working reagent solution. 3.To the sampleJoin18µL premixReagent working solution |
Incubation |
In a microplate shaker800 r/min 37degreeCIncubation30 minutes |
Reading |
Instrument reading,Shade from light/Green light |
* Note: Diluted R1and R2Use the mixture within 30 minutes; it is required toShade from light/Green light;
The reaction system can be scaled proportionally according to experimental requirements.
III. Performance Testing
3.1Example of a complete standard curve:


3.2Performance parameter verification:
LOB:Repeat detection20Next calibratorC1, calculate the signal mean andSD, calculate using the standard curveSignal mean+2timesSDThe corresponding concentration value is the limit of detection (LOD).B.
Matrix |
Blank limit (pg/mL) |
Buffer solution |
1.5 |
Dynamic range (Dynamic range):0 – 300,000 pg/mL.
Linear range (Linear Range):5 – 10,000 pg/mL.
Precision (Precision):
Intra-precision-Precision): In the same experimentOn the microplate, each sample of known low and high concentrations was tested in duplicate.1Zero times, to evaluate the intra-assay precision of this assay.schoolBoth the reference product and the sampleTest using different processes.The coefficient of variation (%CV) for the test results of samples at each concentration was less than 10%, indicating that the method exhibits good reproducibility within the same batch..
Matrix |
Precision |
|
Low concentration |
High concentration |
|
Buffer solution |
3.6% |
20.1% |
Inter-plate precision (Inter-Precision): In 5In the subsequent independent experiments, samples with known low and high concentrations were tested separately. Settings were configured.5The results for each concentration exhibited low coefficients of variation (CV < 10%), indicating that the method demonstrates good reproducibility across different assay batches.
Matrix |
Intermediate precision |
|
Low concentration |
High concentration |
|
Buffer |
8.4% |
5.3% |
Accuracy (Recovery):
Add low, medium, and high concentrations of IL-13standard to different matrices, and repeat the test twice to determine the average recovery rate.
Matrix |
Recovery rate |
||
Low concentration |
Medium concentration |
High concentration |
|
EDTAPlasma |
95.7% |
87.9% |
82.8% |
Heparinized plasma |
105.7% |
94.0% |
87.3% |
Sodium citrate plasma |
104.5% |
105.3% |
96.5% |
Serum |
88.6% |
95.6% |
87.1% |
Culture supernatant |
113.3% |
94.6% |
99.1% |
Dilution linearity (Dilution linearity):
For serum from individuals with high expression, PBMC(5 μg/mL PHA-stimulatedhuman peripheral blood mononuclear cells24 hours) and unstimulated PBMCThe cell supernatants were serially diluted and corrected to evaluate their dilution linearity.


TakeIL-13standard solutionPerform serial dilutions and testing to confirm that samples within the measurement range yield reliable quantitative results, and that no signal suppression occurs.

Test the reference materialNIBSC 94/622, 1 IU ≈ 7.1 ng.
Guidelines
Reagents R1 and R2 should be used protected from light; sample addition and incubation are recommended to be performed under green light or dim lighting (100 LUX).
Each assay requires recalibration, with each concentration level of the calibrator tested in at least duplicate wells, using four-parameter (weighting 1/Y²) curve fitting for calculation.
During incubation, temperature and time must be strictly controlled; it is recommended to cover microplates with a film, and an ELISA reader equipped with the ALPHA function should be used.
Components from different kit lots must not be mixed.